4-5. Operating function
and
function
command
If the instrument is connected with personal com purer through RS-232C
interface. it is possible 10 use personal computer with appropriate
programs. Function
comma
nd designate the program operation to be
carried
out
by instrument,
It
is possible to prepare a set
of
programs
using personal computer to configure the automatic measurement system
having a variety
of
functions.
Th
e functions to be used by using
RS-232C
and
f
unc
tion
com
ma nds are described as follo
ws:
( I) Pan el selling
Panel selling enables the communication with personal co
mp
uter in
both storage or n
on
storage mode. F
or
d
ata
communication, set it
to storage mode to transmit accurate
data
.
(2) Operating function
<D
Wavef
orm
data
transmissi
on
Th
e instrument
has
the following six d
ata
memories.
CHI acquisition memory storing CH I wavef
orm
d
ata
.
CH2 acquisition memory storing
CH
2 wavef
orm
data.
Two sweep trace memories (SAVE A and B)
CH I display memory
CH2 display memory
CH1 and CH2 acquisition memories have each 2000 data l
oad
capacity depending on the sample mod e. CH I and CH2 display
memories two sweep trace memories have 1000
data
l
oad
capacity.
CHI and CH2 display memories
and
two save memories
can
be
transmitted using
per
sonal computer.
Table
4·2
Sampling mode and acquisition
Sample mode
Acquisition memory capacity
R
OLL
1000
NO
RM 2000
EQU IV 2000
Vista de página 93