The second most important measurem
ent
of the synchronous-sweep
oscilloscope is the very measurement of time interval.
Since the divisions unif
or
mly ma rked
on
the CRT screen are all
calibrated to the time bases. the time interval measurement is
prac
ticable,
(I)
Basic Tec
hniq
ue
The basic technique for measuring time interval is described in this
section. In addition. characteristics measurement and variation
technique using this technique will be helpful. if you apply the
following procedure
s:
I. Set up switches in the same
manne
r as that described in 2-2-4.
, Set the TIM E/DIV switch
[n
] in such a manner as
to
make the
\\ aveforrn appear on CRT screen display as large as possible.
Tum the TIM E VARIABLE control [25J fully clockwise until being
clickstopped.
Otherwise you so do. the measured value will be
ina
ccurate. thus
requiring you to exercise due care.
3. Adjust the Vertical POSITION controls [17] and [18] and make the
waveform you wish to measure align with the central horizontal
graticule line.
-+
. Turn the Horizontal Position control [26] and make the left side
of
the waveform correspond to the vertical graci tule line.
5. Count the
numbe
r
of
graticule divisions up to the point you intend
to measure. Horizontal central line is graduated in divisi
on
s
ranging to 0.2 division.